Pointer is not at 0 +/- 3 mmHg
at rest.
Further information
The level of blood pressure is subject to fluctuations even with healthy people. Important thereby
is, that comparable measurements always require the same conditions (rest condition)!
If, in spite of observing all these factors, the fluctuations are more than 15 mmHg, and/or you
hear irregular pulses on several occasions, please consult your doctor.
You must consult your specialist dealer or chemist if there are technical problems with the blood
pressure instrument. Never attempt to repair the instrument yourself!
Any unauthorised opening of the instrument invalidates all guarantee claims!
6. Care and maintenance, recalibration
With proper care and maintenance, this blood pressure measuring device will provide years of
satisfactoryservice.Followthegeneralrulesbelow:
• Donotdrop.
• Neverinflatebeyond300mmHg.
• Donotexposethedevicetoeitherextremetemparatures,humidity,ordirectsunlight.
• Nevercontactthecufffabricwithasharpinstrument,sincethiscouldcausedamage.
• Alwaysdeflatecuffcompletelybeforestorage.
• Donotdismantlemanometerunderanycircumstance.
• Storethewholedeviceinstoragebagprovided,tokeepallthepartsclean.
• Storagetemperaturecondition:20°C to +70 °C at a relative air humidity of 85 % (non-
condensing).
• Wipeoffthemanometerandbulbwithadampcloth.Sterilizationisnotnecessary,since
the parts of manometer should not come into direct contact with the patient‘s body during
measurement.
•
Remove the bladder first, and wipe the closer, bladder and tubes with a damp cloth. The cuff can
be washed with soap and cold water. But do rinse the cuff with clear water and keep it air dry.
10
1. Make sure that the valve is open for zero check.
2. If still more than 3 mmHg deviation, contact your
dealer to recalibrate the manometer.