36
Operating
temp.:
• 10 to 40 °C (50 to 104 °F)
Storage temp. • -20 to 50 °C (-4 to 131 °F)
• 15 - 90 % relative maximum humidity
Weight: • 385 g (including batteries)
Dimensions: • 150 x 100 x 50 mm
Measuring
procedure:
• Oscillometric, corresponding to Korotkoff
Method: • Phase I systolic, Phase V diastolic
Measurement
range:
• 30 - 280 mmHg – blood pressure
• 40 - 200 beats per minute – pulse
Cuff pressure
display:
• Range: 0 - 299 mmHg
• Resolution: 1 mmHg
• Static accuracy: pressure within ± 3 mmHg
• Pulse accuracy: ± 5 % of the readout value
Voltage source: • 4 x 1.5 V Batteries; size AA
• Mains adapter DC 6V, 600 mA (optional)
Reference to
standards:
• Device standard:
Device corresponds to the requirements of
the standard for noninvasive blood pressure
monitor.
EN 1060-1
EN 1060-3
EN 1060-4
IEC 60601-1
IEC 60601-1-2
Electromagnetic
compatibility:
• Device fullls the stipulations of the standard
IEC 60601-1-2.
The stipulations of the EU Directive 93/42/EEC for
Medical Devices Class IIa have been fullled.